Wayne Kerr 4350R | LCR meter, 20 Hz - 500 kHz basic accuracy: 0.1%, USB, LAN, GPIB
Manufacturer number: 4350R
On Request
Wayne Kerr 4350R | LCR meter 4300 series, 20 Hz - 500 kHz, basic accuracy 0.1%, USB, LAN, GPIB.
- Frequency range: 20 Hz… 500 kHz (models up to 1 MHz available)
- Basic accuracy: 0.1%
- Fast measurements: <25 ms
- Measuring functions: C, L, Z, Y, X, B, Rac, Rdc, G, Q, D, Ø
- Single and two test mode
- Up to four parameters that can be displayed
- Interfaces: GPIB / IEEE-488, USB, LAN, RS-232
- Handler Scaleizer module (optional)
- Component binning (optional)
- Dimensions (H x W x D): 104 mm x 322 mm x 285 mm
- Weight: 3 kg
Technical Data
Accuracy: | 0,1 % |
---|---|
Article number: | 1J4350R |
Compensation functions: | Yes |
DC Bias: | internal: +2 V / external: ±40 Vdc |
Dimensions (L x W x H) (mm): | 285 x 322 x 104 |
Displayable measurement parameters: | Cp, Cs, Lp, Ls, Rs, Rp |
Handlerinterface: | Yes |
Interfaces: | USB, LAN, GPIB, RS-232 |
Listsweep: | Yes |
Measurement functions: | Z, L, C, R, D, X, Y, G, B, Ø |
Model: | 4350R |
Resolution: | 5000 mHz |
Special features: | 2 or 4 result parameters two tests simultaneously |
Testadapter: | Yes (Axial, Radial, SMD) |
Testlimits: | Yes |
Warranty (years): | 1 |
Weight (kg): | 3 |
max. Frequency range (MHz): | 0.5 |
min. Frequency range (Hz): | 20 |
Documents & Downloads
Description
4350R | LCR meter, 20 Hz - 500 kHz basic accuracy: 0.1%, USB, LAN, GPIB
Accurate and fast measurements from 20 Hz to 1 MHz (model 43100R) make the 4300 series the ideal choice for many applications in test or production environments.
The two-test mode enables successive measurements with two different instrument parameters. For example, two AC measurements can be performed at different frequencies, or an AC and Rdc measurement can be performed.
An optional component binning module can be purchased. Limits set in the instrument's binning mode allow a variety of components to be sorted, while the choice between isolated and non-isolated modules increases the flexibility of the user interface.
The optional Scaleizer-Handler module offers PASS / FAIL and data synchronization relay outputs for limit-tested measurements. Two output signal configurations are available.
Up to 20 measurement settings can be saved with user-defined names for later retrieval.
All Wayne Kerr Series 4300 LCR meters can be controlled externally via the standard RS232, GPIB, USB or LAN interfaces.
Lieferumfang
- LCR meter
- Measuring cable 1EVA40150
- User Manual
- Power cord
- USB cable
- certificate