Chroma 11210 | Battery cell insulation tester, leakage current measurement Test voltage: up to 1KV (dc) / 50mA

Manufacturer number: 11210

Chroma 11210 | Insulation tester for battery cells, measurement of leakage current and partial discharge, test voltage up to 1 kV DC.

  • Application: Insulation testing of Li-ion batteries as well as various capacitors or insulating material
  • Test voltage: up to 1 kV DC
  • Charging current: max. 50 mA
  • Measuring range for leakage current (LC): 10 pA to 20 mA
  • Measuring range for insulation resistance (IR): 10 Ω to 10 TΩ
  • Partial discharge / flashover detection for the investigation of potential short circuits (option A112100)
  • Logging of the partial discharge and V / I curve shape (option A112101)
  • Integrated DUT contact check
  • Automatic test in the following order: charging - holding - measuring - discharging
  • High speed test: 20 ms / device
  • Color display with 480 x 272 pixels and touch panel for convenient operation
  • Interfaces: handler (control), USB, RS-232, Ethernet
  • Power supply: 90 to 132 V AC or 180 to 264 V AC, 47 to 63 Hz
  • Dimensions: W 320 x H 100 x D 400 mm
  • Weight: 10 kg

Manufacturer-independent

Wide variety of brands from dataTec.

To the brands

Large product portfolio

All the highlights of dataTec Switzerland at a glance.

Download here

Demonstration & Consulting

Sales and technical support close to you.

Start request

Description

11210 | Battery cell insulation tester, leakage current measurement Test voltage: up to 1KV (dc) / 50mA

With the precise battery cell insulation tester 11210 in conjunction with the partial discharge option A112100, the leakage currents of batteries, dry cells, ELDC (Electric Double Layer Capacitors = supercapacitors) and LIC (Lithium Ion Capacitors) can be measured in the millisecond range. In addition to the standard LC / IR measurement (Leakage Current / Insulation Resistance), 11210 has the unique function of detecting partial discharges (PD, partial discharge) and flashovers that can occur inside the insulation material during the testing of the high-voltage insulation. Defective products must be sorted out in the production process before they are filled with the electrolyte. Option A112101 additionally enables the display of the voltage and current curve as well as the recording of PD events with up to 10 images per measurement process.

A sufficiently large charging current is crucial for a high test throughput in the production process. For example, the automated insulation test on the individual DUT can be carried out within 20 ms using the "Charge-Stay-Check-Unload" steps. Leakage currents can be measured from 10 pA to 20 mA in seven current measuring ranges. With the insulation tester 11210, both functions - LC / IR measurement and the detection of partial discharges - are possible simultaneously with one measurement process.

Conventional test systems usually do not differentiate whether a DUT is contacted correctly. Especially with DUTs with a very high internal resistance, a reliable contact check is essential to avoid false positive measurement results. With the integrated function for contact testing, you can exclude contact errors with the 11210 insulation tester within 5 ms before or after the actual insulation test.

Detection and measurement of partial discharge

A constant current (CC mode, constant current) is applied to the DUT during the charging process, the linearity of the voltage increase being checked and an error message being generated in the event of deviations. If the charging voltage is reached, the current is measured in CV mode (Constant Voltage) in order to identify deviations due to leakage current. This way, short circuits in batteries can be reliably detected. Partial discharge detection includes: displaying the partial discharge level and frequency of occurrence, monitoring the partial discharge events during charging and during the dwell phase, programmable partial discharge level limits and logging the partial discharge and V / I waveform (option A112101).

Accessoires (2 products )

Exclusive
Chroma A112101
Partial discharge analyzer card for battery cell insulation tester Chroma 11210
Delivery time upon request
CHF 8,207.00
Exclusive
Chroma A112100
Partial discharge detection card for battery cell insulation tester Chroma 11210
Delivery time upon request
CHF 3,810.00

Options (2 products )

Exclusive
Chroma A112100
Partial discharge detection card for battery cell insulation tester Chroma 11210
Delivery time upon request
CHF 3,810.00
Exclusive
Chroma A112101
Partial discharge analyzer card for battery cell insulation tester Chroma 11210
Delivery time upon request
CHF 8,207.00

About the Manufacturer

Chroma
Chroma ATE is the world's largest supplier of power test instruments and systems, including programmable AC sources and DC power supplies, laboratory power supplies, AC and DC electronic loads, digital meters, multimeters, and automated test systems.

Last Seen Products

Product added to compare
Product Compare added Product Compare